Overview of all controller indications, 1 indications of setting menus – BINDER UF V 500 User Manual
Page 94

UF V (E2+E2.1) 03/2013
page 94/112
Fault description
Possible cause
Required measures
Controller (continued)
Alarm test key cannot actuate the
alarm.
Main power switch (key switch) is
off.
Turn on the key switch (6).
Alarm test key actuates an in-
complete alarm (just message or
just buzzer).
When only the buzzer or only the
alarm is actuated by the alarm
test key, the part, which has not
been actuated is out of order and
must be replaced.
Contact BINDER service.
Automatic door mechanism
UF V (E2): The door cannot be
closed.
The door has been pressed on too
shortly (< 2 sec.) and was opened
immediately after. The automatic
door mechanism was triggered
and now prevents closing the door
again.
Contact BINDER service. Gener-
ally, press on the door for at least
2 seconds until the automatic
door mechanism is activated and
closes the door tightly.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired units must comply with the BINDER quality standards.
17. Overview of all controller indications
17.1 Indications of setting menus
Setting the temperature set point
chap. 6
Operating mode HAND
chap. 8.2
Operating mode USER
chap. 8.3
Locking and unlocking the operating functions’ settings, Operating mode
LOCK
chap. 8.4.1
Password request
chap. 8.3
Password for locking/unlocking
chap. 8.4
Password setting for access to operating mode USER
chap. 8.3
Setting the Safety controller set point (limit value)
chap. 9.1.1
Setting the address
chap. 8.3
Setting the ambient temperature alarm
chap. 8.3