BINDER UF V 500 User Manual
Page 76

UF V (E2+E2.1) 03/2013
page 76/112
Low battery voltage during CO
2
emergency cooling with battery operation
Figure 59: Alarm low battery voltage
during CO
2
emergency cooling with
battery operation
•
Alarm message “
bt.E” immediately
•
LED indication immediately: LED
flashes, LED
flashes
•
Audible alarm: buzzer immediately
•
Switching the zero-voltage relay alarm output.
Actions:
•
Check the battery connections
•
Charge the battery (charging mode: current, approx. 160 mA)
•
If it is defective, replace the battery. Contact BINDER Service
•
Reset the alarm message with the “EXIT“ key.