8 temperature sensor failure – BINDER UF V 500 User Manual
Page 68

UF V (E2+E2.1) 03/2013
page 68/112
10.8 Temperature sensor failure
A sensor fault alarm display takes priority over all other operational
displays and alarm signals on the controller.
Figure 49: Failure of Pt100 temperature sensor for the refrigerating machine. Refrigera-
tion is turned on permanently
Figure 50: Failure of Pt100 temperature sensors for the refrigerating machine
Figure 51: Failure of Pt100 temperature sensor for ambient temperature or air intake of
the 1
st
stage cooling (refrigerating machine)
Figure 52: Failure of Pt100 temperature sensor for safety controller: refrigeration is
turned on permanently
Figure 53: Failure of Pt100 temperature sensor for interior measurement: refrigeration is
turned on permanently
•
Alarm message “
992” or “997” immediately
•
LED indication immediately: LED
flashes
•
Audible alarm: buzzer (intermittent sound) immediately
•
Switching the zero-voltage relay alarm output.
Actions:
•
Please contact BINDER Service.
•
Alarm message “
992” or “998” or “999” immediately
•
LED indication immediately: LED
flashes
•
Audible alarm: buzzer (intermittent sound) immediately
•
Switching the zero-voltage relay alarm output.
•
The refrigerating machine is turned on permanently.
Actions:
•
Please contact BINDER Service.
•
If the temperature rises, i.e. the refrigerating machine is defective (safety controller temperature alarm
and / or tolerance range alarm):
•
Transfer the material to another freezer.
•
Turn off the freezer.
•
If necessary, clean and disinfect the freezer.