3 cleaning and decontamination, 1 cleaning, Ole_link14 – BINDER UF V 500 User Manual
Page 83: Ole_link15, Danger

UF V (E2+E2.1) 03/2013
page 83/112
14.3 Cleaning and decontamination
During operation: Wipe only the outer surfaces with a humid cloth and then dry it thoroughly.
DANGER
Electrical hazard.
Danger of death.
∅
Do NOT spill water or cleaning agents over the inner and outer surfaces
Before cleaning, turn off the unit at the main power switch (key switch)
and disconnect the power plug, thereto pull the locking button on the plug
Completely dry the freezer before turning it on again.
14.3.1 Cleaning
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug, thereto pull
the locking button on the plug (Figure 23).
Wipe the surfaces with a moistened towel.
For a thorough and mild cleaning, we recommend using BINDER Freezy Clean (Art. no. 1002-
0038).
In case of light to normal soiling, we recommend:
•
Cleaning with BINDER Freezy Clean
•
"Final rinsing" with a clean humid wiping cloth
In case of heavier soiling of the interior (except the surrounding ledge made of polycarbonate), we rec-
ommend preliminary cleaning with a standard commercial cleaning detergents free from acid or halides.
Alternatively, you can use the following cleaning agents (apply on a cloth):
Exterior surfaces, housing of
the automatic door mechanism
and controller with controller
panel, interior (stainless steel)
Standard commercial cleaning detergents free from acid or halides.
Alcohol based solutions.
Compartment doors
Standard commercial cleaning detergents free from acid or halides.
Alcohol based solutions max. 10%
Outer door gasket (PVC) and
inner door gasket (silicon)
Alcohol based solutions
Surrounding ledge in the inte-
rior (polycarbonate)
Only use BINDER Freezy Clean (Art. no. 1002-0038)
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
Do NOT use acetone and other organic solvents for cleaning.