2 transport outside a building, Caution – BINDER UF V 500 User Manual
Page 32

UF V (E2+E2.1) 03/2013
page 32/112
To move the freezer through narrow passages (doors, narrow corridors), open the unit door:
Figure 13: UF V 700 with open unit door
For transport outside a building use technical equipment (chap. 3.2.2).
3.2.2 Transport outside a building
Before moving the freezer you can raise the castor locks completely (moving position, chap. 3.5). The
castors are designed only for moving the freezer inside a building (respect the information given in chap.
3.2.1).
We recommend letting down completely the castor locks prior to transport in order to protect the unit
against unintentional rolling.
After operation, please observe the guidelines for temporarily decommissioning the unit (chap. 15.2).
CAUTION
Sliding or tilting the unit.
Damage to the unit.
∅
Do NOT lift or transport the unit using the door, the housing of the automatic door
mechanism or the lower housing.
∅
Do NOT lift the unit by hand
∅
Do NOT transport the unit horizontally.
Transport the unit only in its original packaging.
Secure the freezer with transport straps for transport.
Keep the unit in upright position. Max. angle of inclination during transport: 10°.
Place the shelves on top of each other on the bottom of the interior.
Lift the unit using technical devices (fork lifter) and place it on the transport pallet. Set
the fork lifter laterally or from the rear in the middle of the unit. Make sure to place all
the lateral supports of the unit on the forks (check: the fork protrudes at the opposite
unit side).
Transport units ONLY with the original transport pallet. Set the fork lifter only to the
pallet. Without the pallet the unit is in imminent danger of overturning
Wear suitable shoes (safety shoes).
890 mm
900 mm
140
0 m
m
131
5