Caution – BINDER UF V 500 User Manual
Page 85

UF V (E2+E2.1) 03/2013
page 85/112
Alternatively you can use the following disinfectants (apply on a cloth):
Interior (stainless steel)
Standard commercial surface disinfectants free from acid or halides
(not dripping).
Alcohol based solutions.
Compartment doors
Standard commercial surface disinfectants free from acid or halides
(not dripping).
Alcohol based solutions max. 10%
Outer door gasket (PVC) and
inner door gasket (silicon)
Alcohol based solutions
Surrounding ledge in the interior
(Polycarbonate)
Only use BINDER Freezy Decon (Art. Nr. 1002-0032).
Following use, completely remove any disinfectant with a sterile moistened towel from the
surfaces.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during the
decontamination process.
With every decontamination method, always use adequate personal safety controls.
CAUTION
Danger of corrosion.
Damage to the unit.
∅
Do NOT use acidic or chlorine cleaning detergents.
We recommend using BINDER Freezy Decon (Art. no. 1002-0032) for decontamination.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.