Placing samples in storage in the freezer, Caution – BINDER UF V 500 User Manual
Page 50

UF V (E2+E2.1) 03/2013
page 50/112
7.
Placing samples in storage in the freezer
Before storing valuable samples, conduct a 10-days test run at the desired temperature. This helps to
detect transport damages like capillary cracks. Then you can load the freezer with the precooled sam-
ples.
WARNING: If customer should use a freezer running in non-supervised continuous opera-
tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to
split such specimen or samples and store them in at least two chambers, if this is feasible.
During the equilibration phase of 9 hours after turning on the unit, undefined temperature conditions oc-
cur within the unit. During this phase, do not place any sample materials in the unit
CAUTION
Danger of samples being destroyed.
Load the freezer only after equilibration of temperature.
When placing not precooled samples in storage, temperature rises inside the freezer. We recommend to
successively load the freezer. The more thermal energy you bring in, the longer it takes until reaching the
set-point temperature again.
Do not exceed the maximum load of each compartment and the permitted total load (see chap.18.4).
Always wear protective gloves when opening the freezer and bringing in or removing material.
CAUTION
The inner surfaces become very cold during operation.
Danger of injury by freezing on.
∅
Do NOT directly touch the inner surfaces or the charging material during operation.
∅
AVOID skin contact with the inner surfaces and accessory equipment.
Wear protective gloves when opening the inner doors and during manipulation.