Appendix a – Parr Instrument 6200 User Manual
Page 55
56
6200
Calorimeter Operating Instruction Manual
Communication Controls. Accesses sub-
menus which set the communications
protocols for the printer and balances.
Printer Port (RS232). Accesses sub-menu,
Printer Port Communications. Sets the
communication parameters for the RS232
ports used for the printer port. Standard
options for data bits, parity, stop bits,
handshaking, baud rate and balance type
are provided to match any devices that
might be connected to these ports.
Note:
Default port settings for the 1757 Printer are
8, None, 1, XOn/XOff, 9600. Cable is
A1816E2
Number of Data Bits. Standard
options for data bits. Toggles
between 7 and 8.
Parity. Standard options for parity. Choose
from None, Odd or Even.
Number of Stop Bits. Standard options for
stop bits. Toggles between 1 and 2.
Handshaking. Standard options for
handshaking. Choose from Xon/Xoff,
RTS/CTS and None.
Baud Rate. Standard options for baud rate.
Choose from 19.2K, 9600, 4800, 2400,
1800, 1200, 600, 300, 150, 134.5, 110, and
75.
Printer Type. Toggles between a Parr 1757
and a generic printer. When set for the
1757 Printer, all of the features of this
printer, such as bold printing, will be
activated.
Note:
Serial Ports Pin-Out
9 Pin D
Description
Direction
2 Rx
<
3 Tx
>
4
5 Ground
<>
6
7 RTS
>
8 CTS
<
Both the terminal and balance port are female.
The printer port uses a male connector.
Printer Port Loop Back Test. Used for
factory testing of the printer port.
Communication
Controls Menu
A-12
APPENDIX A