6200

Calorimeter Operating Instruction Manual

Auto Preweigh ID Controls. Accesses
sub-menu, used to automatically assign
Sample ID numbers when a series of
samples are preweighed ahead of the
time they are actually tested.












Automatic Preweigh ID. ON/OFF toggle
for this feature.


Automatic Preweigh ID Prefix.
An entry here will
be used as a prefix for all pre-weigh sample IDs.

Next Automatic Preweigh ID Number.
Shows the
next Sample ID which will be assigned and is
used to enter the beginning Sample ID of any
series

Automatic Preweigh ID Increment.
Establishes
the increment between samples.



Report Width. The column width of the
printer being used can be set to 40 or 80
columns. Select 40 when the 1757 Printer is
used.















Automatic Reporting.
Preliminary reports will
be generated at the conclusion of the test and
final reports will be generated as soon as all of
the thermochemical corrections are available
when this automatic reporting feature is turned
ON. When this is turned OFF, reports will only
be generated through the Report menu.


Automatic
Report Destination. Directs the
reports to the Printer Port or the display.

Individual Reports.
When set to ON, will
generate header information for each report
printed. In the OFF position, only one header
will be printed for a series of tests.

Edit Final Reports.
When set to ON,
enables the user to revise sample weight
and thermochemical corrections.

Recalculate Final Reports. When set to
ON, causes a recalculation of stored final
reports using calibration data and menu
settings currently in the calorimeter.

Use New EE Value in Recalculation.
When set to ON, any recalculation made will
use the most recent EE value in the
calculations. In the OFF position, all
calculations will be made using the EE
value which was effective when the test was
originally run

.

Reporting
Controls
Menu

Data Entry
Controls
Menu
Continued

A-11