Parr Instrument 6200 User Manual

Page 24

background image

25

6200

Calorimeter Operating Instruction Manual

Figure 4- 1

6200 Factory Default Settings

Calorimeter Operation

Operating Mode

Determination

Bomb Installed/EE

1/2400.0

Heater and Pump

OFF



Operating Controls

Method of Operation

Dynamic

Reporting Units

Btu/lb

Use Spiking Correction

OFF

“OTHER” Multiplier

4.1868

Calibrate Touchscreen

LCD Backlight Timeout(s)

1200 S

LCD Contrast

30%

Print Error Messages

ON

Language English


Spike Controls

Use Spiking

OFF

Heat of Combustion of Spike

6318.4

Use Fixed Spike

OFF

Weight of Fixed Spike

0.0

Prompt for Spike before Weight

OFF



Program Info and Control

Date XX/XX/XXXX
Time XX:XX
Software and Hardware Info

Settings Protect

OFF

User/Factory Settings

Feature Key

Bomb Type Select

User Function Setup

Cold Restart



User / Factory Settings

User Setup ID

00000001

Reload Factory Default Settings

Reload User Default Settings

Save User Default Settings



Calibration Data & Control

Calibration Run Limit

10

EE Max Std Deviation

0.0

Heat of Combustion
of Standard

6318.4

Bomb Service Interval

500

Use Bomb

1



Bomb 1 Through 4

EE Value

2400.0

Protected EE Value

OFF



Thermochemical Corrections

Standardization

Fixed Fuse Correction

ON 15.0

Fixed Acid Correction

ON 10.0

Fixed Sulfur Correction

ON 0.0



Determination

Fixed Fuse Correction

ON 15.0

Fixed Acid Correction

ON 10.0

Fixed Sulfur Correction

OFF 0.0

Calculate Net Heat of Combustion OFF


Calculation Factors

Acid Value is Nitric Acid Only

ON

Acid Multiplier

0.0709

Sulfur Value is Percent

ON

Sulfur Multiplier

0.6238

Fuse Multiplier

1.0

Use Offset Correction (ISO)

OFF

Offset Value

0.0

Fixed Hydrogen

OFF 0.0

Hydrogen Multiplier

50.68

Dry Calculation

OFF

Factory Default
Settings

4