Parr Instrument 6200 User Manual
Page 24
25
6200
Calorimeter Operating Instruction Manual
Figure 4- 1
6200 Factory Default Settings
Calorimeter Operation
Operating Mode
Determination
Bomb Installed/EE
1/2400.0
Heater and Pump
OFF
Operating Controls
Method of Operation
Dynamic
Reporting Units
Btu/lb
Use Spiking Correction
OFF
“OTHER” Multiplier
4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s)
1200 S
LCD Contrast
30%
Print Error Messages
ON
Language English
Spike Controls
Use Spiking
OFF
Heat of Combustion of Spike
6318.4
Use Fixed Spike
OFF
Weight of Fixed Spike
0.0
Prompt for Spike before Weight
OFF
Program Info and Control
Date XX/XX/XXXX
Time XX:XX
Software and Hardware Info
Settings Protect
OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User / Factory Settings
User Setup ID
00000001
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Calibration Data & Control
Calibration Run Limit
10
EE Max Std Deviation
0.0
Heat of Combustion
of Standard
6318.4
Bomb Service Interval
500
Use Bomb
1
Bomb 1 Through 4
EE Value
2400.0
Protected EE Value
OFF
Thermochemical Corrections
Standardization
Fixed Fuse Correction
ON 15.0
Fixed Acid Correction
ON 10.0
Fixed Sulfur Correction
ON 0.0
Determination
Fixed Fuse Correction
ON 15.0
Fixed Acid Correction
ON 10.0
Fixed Sulfur Correction
OFF 0.0
Calculate Net Heat of Combustion OFF
Calculation Factors
Acid Value is Nitric Acid Only
ON
Acid Multiplier
0.0709
Sulfur Value is Percent
ON
Sulfur Multiplier
0.6238
Fuse Multiplier
1.0
Use Offset Correction (ISO)
OFF
Offset Value
0.0
Fixed Hydrogen
OFF 0.0
Hydrogen Multiplier
50.68
Dry Calculation
OFF
Factory Default
Settings
4
4-3