Appendix a – Parr Instrument 6200 User Manual
Page 51
52
6200
Calorimeter Operating Instruction Manual
Thermo-
chemical
Corrections
Menu
Print Standardization Runs. Will print all of
the tests that have been incorporated into
the calculated EE value. This will be helpful
in evaluating a series of tests which fail to
produce a satisfactory EE value and relative
standard deviation.
Reset Bomb Fire Count. After bomb
service, press this button to reset the fire
count to zero
Bomb 2. Accesses sub-menu, Bomb 2.
Provides the same controls as described
for Bomb 1.
Bomb 3. Accesses sub-menu, Bomb 3.
Provides the same controls as described for
Bomb 1.
Bomb 4. Accesses sub-menu, Bomb 4.
Provides the same controls as described for
Bomb 1.
Standardization Correction
Fixed Fuse Correction. Displays both the ON
/ OFF of the fixed fuse corrections for
standardization runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set. An appropriate fixed fuse
value is 15 calories.
Fixed Acid Correction. Displays both the
ON/OFF of the fixed acid corrections for
standardization runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set.
Fixed Sulfur Correction. Displays both the
ON/OFF of the fixed sulfur corrections for
standardization runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set. When benzoic acid is used as
the calibrant, a fixed sulfur value of zero
should be used.
Determination Correction
Fixed Fuse Correction. Displays both the
ON/OFF of the fixed fuse corrections for
determination runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set.
Fixed Acid Correction. Displays both the
ON/OFF of the fixed acid corrections for
determination runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set.
Fixed Sulfur Correction. Displays both the
ON/OFF of the fixed sulfur corrections for
determination runs and the value of the
correction. This key toggles the correction
ON/OFF and accesses a sub-menu on which
the value is set.
Note:
When fixed corrections are turned ON, the
value in the specified field will be used in
both the preliminary and final reports. The
calorimeter will not prompt for actual
corrections. If all corrections are fixed, a
preliminary report will not print, rather only a
final report will be generated. If values for
these corrections are entered into these
lines, and the toggle is set to OFF, then the
fixed value will be used in the preliminary
report, but not in the final report.
Calibration
and Data
Controls
Menu
Continued
A-8
APPENDIX A