Troubleshooting

90

PSO-100

Solving Common Problems

No signal is shown in the

application, only samples at the

zero level.

³

Power is too low.

³

Wavelength is not

appropriate.

³

Dirty contacts.

³

Increase signal

power.

³

Use the

1520-1565 nm

wavelength

region.

³

Clean the input

contacts. If you

send in a lot of

power and the

contacts are dirty,

the input contact

may be damaged

and must be

repaired.

Windows indocates that it

found new hardware when you

insert the USB cable

³

You are using a

new USB port

(Windows XP).

³

The USB driver is

not installed.

³

Use the correct

USB port (the one

you used when

the USB driver

was installed) or

install the driver

on the new port.

³

Install the USB

driver.

Problem

Cause

Possible Solution