Setting up and Using the Optical Sampling Oscilloscope

Optical Sampling Oscilloscope

33

Managing SoftSync Input

Managing SoftSync Input

The SoftSync inputs section the bottom left is used for input of parameters
important for the software synchronization algorithm to correctly
synchronize the data.

In order to obtain the correct time scale, an estimate of the signal
frequency should be inserted since there is no other way for the software
to know the bit rate. The signal frequency should be accurate within the
correct 8 MHz (sampling frequency divided by two) interval, according to
the figure below.

Ensuring that the correct values are entered ensures appropriate
measurements.

Signal

frequency

Number of eye periods to display

on-screen (available in eye diagram mode)

Data symbol rate (for example,

1/4 for a signal frequency of

40 GHz and a clock frequency of

10 GHz).

Pattern length of the input signal
(available in pattern mode)

...

...

Signal frequency

8 MHz

Even Intervals

Odd intervals