Troubleshooting

92

PSO-100

Solving Common Problems

The sampled data looks very

noisy, jittery or distorted. The

software does not seem to be

able to synchronize the data.

³

The input signal is

too low.

³

The signal has

very poor OSNR

or suffers from

dispersion.

³

The waveform

format is

incorrect.

³

There are

frequency holes.

³

The data signal

suffers from large

phase noise.

³

The signal

wavelength is in

the 1567-1570 nm

span.

³

The internal

sampling pulse

source has not

locked yet.

³

Increase the

signal power

³

Try to use the

external clock

³

Set the

appropriate

waveform format

or use the

external clock.

³

Click clear.

³

Change the phase

tracking

algorithm

(Utilities >
Advanced

menu), or use the

external clock in

adaptive mode.

³

Use

1525-1565 nm.

³

Turn off the

optical input

signal. Wait and

let the unit warm

up until it locks.

My OTDM data jumps around

on the screen.

It is difficult to

synchronize a

sub-optimally

multiplexed data

signal.

Use the external

clock with a subrate

clock.

Problem

Cause

Possible Solution