Troubleshooting

Optical Sampling Oscilloscope

93

Solving Common Problems

My short RZ pulses look like an

eye diagram.

It is difficult to

correctly synchronize

very low duty cycle

pulses <1 %.

Start and stop the

sampling to

resynchronize the

pulses, until it looks

like short pulses. If it

does not work, use

the external clock.

The eye diagram looks jittery

even if you use an external

clock

The clock signal may

suffer from large

phase noise.

Change the phase

tracking algorithm

(Utilities >
Advanced menu), or

use the adaptive

clock mode.

The DC level is not zero even if

the input signal is off.

The DC level drifts

slightly with

temperature.

Adjust the zero level

from the Utilities

menu.

The samples are unevenly

distributed or appear at discrete

points in the signal. When you

start and stop the sampling the

data may be incorrectly

synchronized.

Frequency holes.

Wait and see if it

disappears. Increase

the number of layers

and/or the number of

samples/layer.
Tune the signal

frequency a few

100 Hz if possible.

The application crashes. Error

messages appear in the DOS

window.

Software bug.

Click Start again. If

that does not help,

restart the

application. Contact

EXFO.

Problem

Cause

Possible Solution