Optical Sampling Oscilloscope

87

6

Troubleshooting

Solving Common Problems

Problem

Cause

Possible Solution

Nothing happens when you

start sampling, but you can use

the Start/Stop button.

³

The samples are

outside the

sampling screen.

³

The input power

is too high.

³

Zoom out or pan

to find the

samples.

³

Reduce input

power.

The application freezes and you

cannot use the Start/Stop

button.

³

The unit was

turned off while

the application

was sampling.

³

Turn on the unit

again, then restart

the application on

the host

computer.

The application does not start

at all

³

The MCR is not

installed

³

The system path

is incorrect

³

The MCR version

is incorrect

³

Install the MCR

using the

MCRinstaller

³

See Installing the
Matlab Runtime,
USB Drivers and
Optical Sampling
Oscilloscope
Software
on

page 12.

³

When receiving a

new software

update from

EXFO, make sure

that you use the

correct MCR

version.