beautypg.com

Input:step, Input:step:catalog – EXFO FVA-3150 Variable Attenuator User Manual

Page 99

background image

Remote Control Commands

Variable Attenuator

93

Specific commands-Quick Reference Command Tree

INPut:STEP?

Description This command is used to obtain the current step size between

two attenuation settings.

Syntax

INP:STEP?

Response

The current step size in the format “0.02”.

Example

INP:STEP?

INPut:STEP:CATAlog?

Description This query returns the list of available step sizes between two

attenuation settings. Use this command to determine the

maximum resolution of the attenuation or reflection setting.

Syntax

INP:STEP:CATA?

Response

The list of available steps. The steps are divided by “;” and the

units are dB. For example: 0.01;0.02;0.05;0.1;0.2;

0.5;1;2;5;10;20;50

Example

INP:STEP:CATA?