beautypg.com

Output[:state – EXFO FVA-3150 Variable Attenuator User Manual

Page 102

background image

Remote Control Commands

96

FVA-3150

Specific commands-Quick Reference Command Tree

OUTPut[:STATe]

Description This command controls the shutter, which can block optical

continuity.

Syntax

OUTP[:STAT]

Parameters The parameter can be

“0” or “OFF” to optically enable the shutter
“1” or “ON” to optically disable the shutter

Example

OUTP:STAT ON

Note

This command cannot be issued while a program is running.

OUTPut[:STATe]?

Description This query returns the state of the shutter.

Syntax

OUTP[:STAT]?

Response

“0” if the shutter is optically enabled
“1” if the shutter is optically disabled

Example

OUTP:STAT?