beautypg.com

Program:selected:attenuation, Program:selected:delay – EXFO FVA-3150 Variable Attenuator User Manual

Page 103

background image

Remote Control Commands

Variable Attenuator

97

Specific commands-Quick Reference Command Tree

PROGram:SELected:ATTenuation

Description This command changes the attenuation parameter for the

attenuation program.

Syntax

PROG:SEL:ATT[]

Parameters The parameter is the difference in attenuation

between each step in the program (maximum 99.999 dB). The

units are optional.

Example

PROG:SEL:ATT 3

Note

This command cannot be issued while a program is running.

PROGram:SELected:DELAy

Description This command changes the delay parameter for the attenuation

program.

Syntax

PROG:SEL:DELA

Parameters The parameter is the delay preceding the program

start. This value has the following format: “999,59,59”

(hours,minutes,seconds).

Example

PROG:SEL:DELA 000,05,30

Note