Remote Control Commands

Variable Attenuator

91

Specific commands-Quick Reference Command Tree

INPut:OFFSet

Description This command stores a reference value for the current

wavelength and sets the FVA-3150 to Reference mode. In

Reference mode, the displayed attenuation is relative to this

reference value. A different reference value must be stored for

every wavelength. The reference values remain in memory

until they are overwritten.

Syntax

INP:OFFS[DB]

Parameters The parameter must be a valid attenuation value

followed by the (optional) units (dB).

Example

INP:OFFS 99.99 DB

Note

This command cannot be issued while a program is running.

INPut:OFFSet:DISPlay

Description This command sets the reference value (for the current

wavelength) to the current absolute attenuation setting and sets

the FVA-3150 to Reference mode. In Reference mode, the

displayed attenuation is relative to the reference. A different

reference value must be stored for every wavelength. The
reference values remain in memory until they are overwritten.

Syntax

INP:OFFS:DISP

Parameters N/A

Example

INP:OFFS:DISP

Note

This command cannot be issued while a program is running.