Remote Control Commands

94

FVA-3150

Specific commands-Quick Reference Command Tree

INPut:WAVElength

Description This command selects a specific calibrated wavelength. The

wavelength range depends on the type and configuration of the

FVA-3150.

Syntax

INP:WAVE[]

Parameters The parameter is the actual wavelength in

the format “9999.9 NM”. The units are optional.

Example

INP:WAVE 1310 NM

Note

This command cannot be issued while a program is running.

The INP:WAVE command is equivalent to the WVL command.

Both give the same result.

INPut:WAVElength?

Description This query returns a value indicating the currently selected

wavelength, the minimum available wavelength, or the

maximum available wavelength.

Syntax

INP:WAVE?[]

Parameters The parameter is optional and can be

“MIN” to request the minimum wavelength
“MAX” to request the maximum wavelength

Response

The wavelength in the format “9999.9”. The units are nm.

Example

INP:WAVE? MAX or
INP:WAVE?

Note

The INP:WAVE? command is equivalent to the WVL? command.

Both give the same result.