Remote Control Commands

100

FVA-3150

Specific commands-Quick Reference Command Tree

WaVeLength

Description This command selects a specific calibrated wavelength. The

wavelength range depends on the type and configuration of the

FVA-3150.

Syntax

WVL[]

Parameters The parameter is the actual wavelength in the

format “9999.9[NM]”. The units are optional.

Example

WVL 1310 NM

Note

This command cannot be issued while a program is running.

The WVL command is equivalent to the INP:WAVE command.

Both give the same result.

WaVeLength?

Description This query returns a value indicating the currently selected

wavelength, the minimum available wavelength, or the

maximum available wavelength.

Syntax

WVL?[]

Parameters The parameter is optional and can be

“MIN” to request the minimum wavelength
“MAX” to request the maximum wavelength

Response

The wavelength in the format “9999.9”. The units are nm.

Example

WVL? MAX or
WVL?

Note

The WVL? command is equivalent to the INP:WAVE? command.