Remote Control Commands

Variable Attenuator

83

Specific commands-Quick Reference Command Tree

ATTenuation?

Description This query returns a value corresponding to the CAL parameter

value added to either the current attenuation setting, the

minimum attenuation, or the maximum attenuation setting.

Syntax

ATT? []

Parameters The parameter is optional and can be

“MIN” to obtain the minimum setting for the current wavelength
“MAX” to obtain the maximum setting for the current

wavelength

Response

The attenuation value in the format “99.999”. The units are dB.

Note that, depending on the current measurement mode, the

response will include different parameters:
In Absolute mode:

In Reference mode:

In Offset mode:

Example

ATT? MIN

Note

The ATT? command is not equivalent to the INP:ATT?

command.

response

actual attenuator general offset (CAL?)

+

=

response

actual

attenuation

reference value

INP:OFFS?

general offset

(CAL?)

+

=

response

actual

attenuation

wavelength-specific

offset

general offset

+

+

=