Using Your Broadband Source in an Automated Test Environment

Broadband Source

47

SCPI Command Structure

There are also several conventions regarding command syntax:

Spelling errors will cancel the command or query.

Commands and queries are not case-sensitive. You can type your
program messages using either lower-case or upper-case letters.

The command or query can be written using only the three- or
four-letter shortcuts, only full words, or a combination of both.

The example below shows the long and the short forms of a same
query.

For readability reasons, you can use extra spaces in your program
messages but they won’t be taken into account. For more information,
see Message Management on page 35.

;

Mandatory to separate the different commands of a program

message when more than one command is sent at a time. In this

case, it is called .

Also used to separate responses when multiple queries were sent in

a single program message. In this case, it is called MESSAGE UNIT SEPARATOR>.

,

Mandatory to separate parameters in a command or a query. In this

case, it is called .

Also used to separate the various responses from a query. In this

case, it is called .

Item

Meaning

:SYSTem:ERRor?

:SYST:ERR?
:syst:err?

Long form

Short form (small words
represented by the capital letters
of the long form)