Attention (OUT_MODE): If control is running, it is only possible to switch either from 1 to 2, or
from 2 to 3, or from 3 to 2. The set vacuum is adopted in each case.
Command
Operation
Response
Description
IN_SP_4
delay
XX:XX h:m
hours:minutes (00:00 = Off)
IN_SP_5
switch off
pressure
XXXX mbar/Torr/hPa
”Maximum” for ”Vac control”,
”Minimum” for ”Pump down”)
unit according to preselections
IN_SP_6
runtime
XX:XX h:m
process runtime (hours:minutes)
IN_SP_P1y
time
XX:XX:XX h:m:s
time in program step y (0......9)
(hours:minutes:seconds)
IN_SP_P2y
pressure
XXXX mbar/Torr/hPa
pressure in program step y (0......9)
unit according to preselections
IN_SP_P3y
venting valve
0
1
no venting valve in program step y
(0......9)
venting valve in program step y
(0.....9)
IN_SP_P4y
Step
0
1
no ”Step” in program step y (0.....9)
”Step” in program step y (0......9)
IN_SP_P5y
Auto
0
1
no ”Auto” in program step y (0.....9)
”Auto” in program step y (0......9)
IN_PV_Sx
current pressure
of pressure trans-
ducer x
XXXX mbar/hPa/Torr
pressure of pressure transducer x
(order of numbering according to
display in ”Sensors” menu)
IN_VER
version
CVC 3000 VX.XX
software version
Write commands ”CVC 3000”
Command
Operation
Parameter
Description
OUT_MODE
function
0
1
2
3
30
31
32
4
VACUU•LAN
Pump down
Vac control
Auto mode
optional: sensitivity: low
optional: sensitivity: normal
optional: sensitivity: high
Program
OUT_CFG
configuration
(bus monitoring)
yXXX
X0XX
X1XX
X2XX
XX0X
XX1X
XXX0
XXX1
y: 0......D: language
+
(hexadecimal), see
”Read commands CVC 3000”
pressure unit mbar
pressure unit Torr
pressure unit hPa
Autostart off
Autostart on
acoustic signal off
acoustic signal on