Attention: Never use a pointed or sharp-edged tool to

clean the pressure transducer.

Never touch the ceramic diaphragm of the pressure

transducer with hard objects.

- Fill the measurement chamber with a solvent (e.g.,

benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of sol-

vents!

- Drain the solvent and dispose of in accordance with

regulations. Repeat cleaning if necessary.

- Rinse the measurement chamber several times with

alcohol in order to remove all solvent residues.

- Allow the pressure transducer to dry.

- Readjust the pressure transducer if necessary.

NOTICE

Readjustment of the controller CVC 3000

See section “Readjustment of CVC 3000”, pg. 78.

Cleaning the pressure transducer