beautypg.com

Pump down function – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 57

background image

page 57 of 129

Pump down function

Continuous pumping with pressure and time settings

• Operation on demand of a speed controlled NT VARIO / VARIO-B

pump

Preselections
+

Use the selection knob to select the parameters. All parameters can

be altered even while operation control is running.

+

Speed: Preselection of the motor speed for pump down. The selec-

tion ”HI” effects the maximum speed and best ultimate vacuum of the

pump (with automatic speed reduction at ultimate vacuum).

The ”Speed” is adjustable in a range of 1-100% and to ”HI”.

+

Minimum: The controller switches the pump off once the preset value

for ”Minimum” has been reached.

Minimum” is adjustable in a range of 1-795 Torr (1-1060 mbar) or can

be set to ”Off”.

A preset ”Duration” (process time) has no effect if the process is

stopped due to a preset ”Minimum” before ”Duration” is reached.

+

Delay: Delay” determines the time the coolant valve remains open

after the process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately when the

process stops.).

+

Duration: Duration” determines the total process time since control

start.

The process time is adjustable between 1-1440 minutes (24 h) or can

be set to ”Off”. ”Off” indicates that no endpoint for pump down is deter-

mined.

If a ”Duration” is preset, the controller switches off the pump when the

preset process time is reached, even if a preset ”Minimum” is still not

reached.