VACUUBRAND PC 3012 VARIO plastic housing User Manual
Page 42
page 42 of 129
Catchpots:
The catchpot at the inlet protects against droplets and par-
ticles from entering the pump.
+
Enhances lifetimes of diaphragms and valves.
+
Improves vacuum performance in applications with
condensable vapors.
PC 300x VARIO:
Both catchpots are coated with a protective
layer to protect against shattering in case of
breakage or implosion.
➨
Assemble the catchpots at the inlet and at
the outlet using joint clips.
PC 301x VARIO:
➨
Assemble the inlet catchpot (18) with cen-
tring ring and clamping ring at the inlet of
the pump.
➨
Assemble the collecting flask (20) at the
exhaust waste vapor condenser using the
joint clip.
➨
In addition to connecting the pump inlet to
the application, install a vacuum connection
(hose) between the pressure transducer
(measurement connection at the rear of the
CVC 3000) and the vacuum system.
Exhaust waste vapor condenser:
➨
Assemble the hose nozzles for coolant inlet
(15) and coolant outlet (16) tubing and for
the gas outlet (8) at the exhaust waste va-
por condenser.
The
exhaust waste vapor condenser en-
ables an efficient condensation of the pumped
vapors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass
catchpot at
outlet (14)
catchpot
at inlet (14)
(18)
(19)
(15)
(16)
PC 300x VARIO
outlet (gas!) (8)
(15)
(16)
outlet (gas!) (8)