beautypg.com

Gas inlet temperatures – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 27

background image

page 27 of 129

Gas inlet temperatures

* Ultimate vacuum at setting ”Pump down” at speed ”HI”

** Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 1500rpm and ulti-

mate vacuum with exhaust tube at outlet.

Type

ME 4C NT

VARIO

MZ 2C NT

VARIO

MD 4C NT

VARIO

Maximum pumping speed

50/60 Hz (ISO 21360)

cfm

(m

3

/h)

2.9

(4.9)

1.66

(2.8)

2.7

(4.6)

Ultimate vacuum (abso-

lute) without gas ballast*

Torr

(mbar)

52.5

(70)

5.3

(7)

1.1

(1.5)

Ultimate vacuum (abso-

lute) with gas ballast*

Torr

(mbar)

-

-

9

(12)

2.3

(3)

Maximum rated current at:

100-120 V~ 50/60 Hz

200-230 V~ 50/60 Hz

A

A

3.2

1.4

3.2

1.4

6.3

2.5

Device fuse

100-120V: slow blow fuse 8A

200-230V: slow blow fuse 4A

Degree of protection IEC 529

IP 20

Inlet

hose nozzle for tubing I.D. 3/8“

(hose nozzle DN 10 mm)

Outlet

silencer or

-

hose nozzle for tubing I.D. 3/8“

(hose nozzle DN 10 mm)

A-weighted emission

sound pressure level**

(uncertainty K

pA

: 3 dB(A))

dB(A)

45

43

Dimensions L x W x H

approx.

in

(mm)

10.0 x 9.57 x

9.65

(254 x 243 x

245)

9.57 x 9.57 x

9.65

(243 x 243 x

245)

12.8 x 9.57 x

9.65

(325 x 243 x

245)

Weight approx.

lbs. (kg) 30.4 (13.8)

30.4 (13.8)

35.9 (16.3)

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)