Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Catchpot or collecting flask at outlet:

Remove joint clip. Remove catchpot

or collecting flask and

drain condensate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to restore at-

mospheric pressure in the catchpot before attempting

removal. Remove joint clip. Remove catchpot and drain

condensate.

Catchpot at inlet (PC 301x VARIO):

Admit air or inert gas (via the pump inlet) to restore atmo-

spheric pressure in the catchpot before attempting remov-

al. Open drain screw (19) and drain condensate. Close

drain screw.

Reattach drained catchpots.

+

Important: Comply with regulations when disposing

of solvents/condensates. Recycle if possible; purify if

contaminated.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

NOTICE

NOTICE