page 88 of 129

Command

Operation

Response

Description

IN_STAT

status process

control

0XXXXX
1XXXXX
X0XXXX
X1XXXX
XX0XXX
XX1XXX
XXX0XX
XXX1XX
XXXX0X
XXXX1X
XXXX2X
XXXX3X
XXXX4X
XXXX5X
XXXXX0
XXXXX1

XXXXX2

XXXXX3

pump off
pump on
in-line valve closed
in-line valve open
coolant valve closed
coolant valve open
venting valve closed
venting valve open
VACUU•LAN
Pump down
Vac control
Auto mode
Program
measuring device
control off
pump down -
determining boiling point
set vacuum reached -
boiling pressure found
current pressure below set vacuum
- automatic switch-off

IN_ERR

fault status

0XXXXXXXX
1XXXXXXXX
X0XXXXXXX
X1XXXXXXX
XX0XXXXXX
XX1XXXXXX
XXX0XXXXX
XXX1XXXXX
XXXX0XXXX
XXXX1XXXX
XXXXX0XXX
XXXXX1XXX
XXXXXX0XX
XXXXXX1XX
XXXXXXX0X
XXXXXXX1X
XXXXXXXX0
XXXXXXXX1

no fault at pump
fault at pump
no fault at in-line valve
fault at in-line valve
no fault at coolant valve
fault at coolant valve
no fault at venting valve
fault at venting valve
no overpressure
overpressure
no fault at pressure transducer
fault at pressure transducer
catchpot not full
catchpot full
no external fault
external fault
last interface command correct
last interface command incorrect

IN_SP_1

set vacuum

XXXX mbar/Torr/hPa

unit according to preselections

IN_SP_2

maximum speed

XXX%

speed in % (1-100% or ”HI”)

IN_SP_3

switching

pressure

XXXX mbar/Torr/hPa

switching pressure for VACUU•LAN
or two point control;

unit according to preselections