Read commands CVC 2000

Command

Operation

Response

Description

IN_PV_1

current pressure

XXXX mbar/

Torr/hPa

unit according to preselections

IN_PV_2

current frequency

XX.X Hz

pump speed

IN_CFG

device set

preselections

0XXXX

1XXXX

2XXXX

3XXXX

X0XXX

X1XXX

XX0XX

XX1XX

XXX0X

XXX1X

XXXX0

XXXX1

VACUU•LAN

continuous pumping

vacuum control without automatic

vacuum control with automatic

no coolant valve

coolant valve

no venting valve

venting valve

no automatic switch off

automatic switch off

remote operation off

remote operation on

IN_ERR

error code

1XXX

X1XX

XX1X

XXX1

fault at pump electronics

overpressure

maloperation mode pressure transducer

last command to interface incorrect

IN_STAT

status of

process control

0XXX

1XXX

X0XX

X1XX

XX00

XX01

XX02

XX03

XX10

XX11

XX20

XX21

XX22

XX23

XX30

XX31

XX32

XX33

coolant valve closed

coolant valve open

venting valve closed

venting valve open

VACUU•LAN: inactive

VACUU•LAN: pumping down,

current pressure > selected pressure

VACUU•LAN: pumping down, time for

automatic switching off is running

VACUU•LAN: system is switched off

continuous pumping: not active

continuous pumping; active

vacuum control: not active

vacuum control:

current pressure above set vacuum

vacuum control: current pressure

equals set vacuum (

±1 Torr/mbar)

vacuum control:

current pressure below set vacuum

automatic: not active

automatic: determining boiling point

automatic: adjusting boiling point

automatic: system is switched off