beautypg.com

Notice – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 47

background image

page 47 of 129

+

Never block the gas outlet ((8) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose must always be

unobstructed and without back-pressure to enable an

unhindered discharge of gases.

+

Check the overpressure safety relief device (17) at

the exhaust waste vapor condenser (13) regularly; re-

place if necessary. Check especially for deterioration,

coalescence and cracks.

• Ensure that the

coolant outlet hose is always free

and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• We strongly recommend installing an optional coolant

valve (see “Accessories”, pg. 92)

in the supply line

of the exhaust vapor condenser to save water and re-

duce the risk of water spill.

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or kinked coolant hoses).

In case of

condensation: Check the liquid level in both

catchpots (14) or in the collecting flask (20) during op-

eration. Check the liquid level regularly. Do not allow the

catchpots or the collecting flask to overfill. Drain catch-

pots or collecting flask in time to avoid overflow. Install a

level sensor (see “Accessories”, pg. 92) for monitoring,

if necessary.

The maximum liquid level is at approximately 80% of the

total filling level to avoid problems when removing the

catchpots or the collecting flask.

Permissible range of coolant temperature at the exhaust

waste vapor condenser:

5 °F to 68 °F (-15°C to +20°C)

NOTICE