beautypg.com

Vac control function – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 59

background image

page 59 of 129

Vac Control function

Vacuum control to a preset vacuum value

• Operation on demand of a speed controlled pump (NT VARIO /

VARIO-B)

Preselections
+

Use the selection knob to select the parameters. All parameters can

be altered even while operation control is running.

+

Set vacuum: The ”Set vacuum” is the set point for vacuum control

with pinpoint precision for NT VARIO / VARIO-B pumps.

The ”Set vacuum” is adjustable in a range of 1-795 Torr (1-1060 mbar)

or can be set to ”Turbo”. The selection ”Turbo” leads to the optimum

backing pressure for a turbomolecular pump.

+

Speed: The maximum motor speed can be preselected (to control the

pumping speed).

The ”Speed” is adjustable in a range of 1-100% or can be set to ”HI”.

The selection ”HI” (recommended) provides the optimum pumping

speed of the pump and therefore an optimum ultimate vacuum.

+

Maximum: An upper pressure limit can be preselected. The pump

switches off if the pressure limit is exceeded, e.g., at the end of suc-

tions or filtrations. The pressure limit is only active once the pressure

has gone below ”Maximum”.

The ”Maximum” is adjustable in a range of 794-1Torr (1059 - 1 mbar)

(at the least 1 Torr (mbar) higher than the ”Set vacuum”) and to ”Off”.

”Off” means that no ”Maximum” value is preset.

+

Delay: Delay” determines the time the coolant valve remains open

after the process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately when the

process stops.).

+

Duration: Duration” determines the total process time since control

start.