Maintenance and repair, Notice – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 22

background image

page 22 of 129

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Press the START/

STOP key at the controller to confirm the error message.

Switch off the pump and disconnect from the power source.

Identify and eliminate the cause of failure. Wait approx.

five minutes before restarting the pump.

Maintenance and repair

In order to comply with laws (occupational, health and

safety regulations, safety at work law and regulations for

environmental protection) vacuum pumps, components

and measuring instruments can only be returned when

certain procedures (see section “Notes on return to the

factory”, pg. 122) are followed.

Take advantage of our service seminars, which put special

focus on the maintenance and repair of vacuum pumps.

For details and for the online ”Instructions for repair” man-

ual see www.vacuubrand.com.

In normal use, the lifetime of the diaphragms and valves is

typically 15,000 operating hours. Bearings have a typical

durability of 40000 h.

Ensure that the pump cannot be operated acciden-

tally. Never operate the pump if covers or other

parts of the pump are disassembled.


Switch off the pump. Disconnect the electrical

power cord and wait two minutes before starting

maintenance to allow the capacitors to discharge.

Note: The pump may be contaminated with process

chemicals, which have been pumped during operation.

Ensure that the pump is completely decontaminated

before maintenance commences.

NOTICE