Configuration – VACUUBRAND PC 3012 VARIO plastic housing User Manual
Page 73
page 73 of 129
Configuration
In the ”Configuration” menu the device parameters are preselected.
After 20 seconds without action the function ”Configuration” and its sub-
menus (except submenu ”Sensors”) are quit automatically without stor-
ing any possibly changed parameter.
+
Adjustment: Adjustment of the pressure transducer under vacuum
and/or at atmospheric pressure, see also section “Readjustment of
Adjustment to atmospheric pressure is carried out at an absolute pres-
sure value between 795 - 525 Torr (1060 - 700 mbar). This is espe-
cially helpful in high elevation laboratories. Adjustment under vacuum
may be done at an absolute vacuum value between 0 - 15 Torr (0 - 20
mbar). In the range between 15 to 525 Torr (20 to 700 mbar) no adjust-
ment is possible; ---- Torr is displayed.
+
RS-232: Configuration of the interface, setting of parameters and com-
mands, see section ”Interface”.
Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,
”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and
remote on ”On” or ”Off”.
+
Sensors: Selection of the pressure transducer to be controlled (maxi-
mum 4 external sensors).
+
Display: Selection of the device parameters ”Brightness” between 0
- 100%, ”Contrast” between 0 - 100%, ”Sound” ”On” or ”Off”, ”Units”
”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”, ”French”,
”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”,
”Russian”, ”Polish”, ”Dutch”, ”Japanese”, ”Finnish”.
+
Autostart: If ”Autostart” is set to ”On” the controller restarts a running
process automatically after a mains failure. If this is unwanted, set
”Autostart” to ”Off”.
Attention: If ”Autostart” is preselected, the process starts
immediately after power failure without pressing any fur-
Preselections
+
Use the selection knob to select the parameters.