LABEC M-MIA-A User Manual
Page 43

Ultrasonic milk analyser
Serves to adjust the contrast of the display. The parameter may be changed 
in two modes: 
- fast – for rough adjustment; 
- fine – for precise adjustment. 
 
After starting this function the displays shows the following: 
 
 
 LCD Setup: 
LCD Setup 
Contrast fast 
Contrast fine 
B. Light fast 
------------------------------
B.Light fine 
Exit 
 
 
7.2.3.4.Recalibrate. 
Serves for changing definite calibration. Methods are described in point 6.4. 
 
7.2.3.5. Edit FrPoint 
Through this menu you have the possibility to enter the basic freezing point 
separately for each calibration. For more information, see Appendix Freezing 
point.
After choosing the menu the following is displayed:
 
Edit FrPoints 
FrPoint Calibr1 
FrPoint Calibr2 
FrPoint Calibr3 
-----------------------------
 Exit 
 
 
After choosing freezing point for the calibration, the following is displayed: 
 
FrPoint Calibrx
-0.fff
- OK +
 
Where: 
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)