LABEC M-MIA-A User Manual
Page 39

Ultrasonic milk analyser
After 5th measurement completion automatically appears the menu:
Recal 1
Put sample water
Which reminds for 5-times water measurement. There is no need to enter
values after placing the sample-holder. Measurement starts directly.
After the 5
th
measurement appears the menu:
Recalibrated
Analyzer ready
This means that the calibration was completed successfully and the analyzer
is recalibrated for cow milk, marked as “Calibr 1”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.
If during work the sample’s temperature exceeds the temperature range 15-
25° С the following message appears:
Temperature out of
range
Wait till the end of the measurement. When the message appears:
Put sample again
Put sample with temperature in the temperature range and continue the
measurement till completion of 5-time measurement.
Operation manual
MRC.VER.01-02.10
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)