7. STARTING THE DEVICE IN A SERVICE TEST/SETUP
OPERATIONAL MODE. MENUS DESIGNATION
7.1. Starting the device in a service Test/Setup operational mode.
In order to start the Setup of the device the operator has to press the button
Enter and without releasing it to switch on the power supply of the device, to
wait for the starting identification messages and to release the button after
the following message appears on the display:
Release button
to start setup
After releasing the button on the display is shown:
MA Setup
Followed by possible to be entered by the operator menus:
Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Exit
You may move in the menus by using buttons “up” ▲ and “down” ▼.
If by pressing the button Enter you choose a menu, each menu offers new
points/submenus. When Exit is chosen the device leaves the Setup mode
and returns to normal work.
7.2. Menus Function:
7.2.1. Special modes.