LABEC M-MIA-A User Manual
Page 38
Ultrasonic milk analyser
and it is put in the recess of the device. Starts the measurement. The sample
is sucked. Appears the following menu:
Recallibrate 1
Put sample high 5
times
Samp T=16.8
After the sample is measured, appears the following menu:
Recalibrate 1
Put sample High
Cal meas=1/5
Follow the procedure till the 5th measurement.
After 5th measurement completion automatically appears the menu:
Cal 1 Samp Low
Fat = f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit
Which reminds to enter the values for the sample with low FAT. Enter the
values analogically to the procedure, described in the previous sample. After
entering the last parameter position the cursor on “Exit”, press “Enter” and
the following menu will appear:
Recal 1
Put sample Low 5
times
Make 5 times measurement of the low FAT sample.
Operation manual
MRC.VER.01-02.10
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)