Ultrasonic milk analyser

Follows rinsing with water:

The operator has to put a glass with water for rinsing the device.


Put filled with

Water glass

and press Enter

to Continue


After rinsing the display shows the following:

End of cleaning


After 2 seconds the following appears on the display:

Analyzer ready


After this the analyzer is ready for normal measurement or to be switched off
from the power supply.

4.2. Preparation of cleaning solution

4.2.1. Preparation of alkaline cleaning solution
Preparation of 1 % alkaline solution of Lactodaily for circulation cleaning in
the milk analyzer:
1. Take the package 100 g concentrated powder chemical Lactodaily
2. Carefully cut the upper end, paying attention not to spill it.
3. In appropriate vessel (for example bucket) pour 1 l water.
4. Add the powder and then again water up to 10 l.
Then follow the instruction for milk analyzer cleaning.

4.2.2. Preparation of acidic cleaning solution

Preparation of 1 % acidic solution of Lactoweekly for circulation cleaning in
the milk analyzer:
1. Take the package 100 g concentrated powder chemical Lactoweekly
2. Carefully cut the upper end, paying attention not to spill it.