LABEC M-MIA-A User Manual
Page 36
![background image](/manuals/697908/36/background.png)
Ultrasonic milk analyser
6.4.1.3. Analogically, position on Recalibrate and press the button Enter.
6.4.2. Making recalibration
6.4.2.1. Determining the calibration mode
After starting Recalibrate, the display shows the following:
Recalibrate
Calibration 1
Calibration 2
Calibration 3
-----------------------------
Edit samp’s 1
Edit samp’s 2
Edit samp’s 3
Edit FrPoints
-----------------------------
Exit
Using the buttons “up” ▲ and “down” ▼ position on the corresponding
calibration (for example if you need to recalibrate cow milk, which is assigned
in the milk analyser as Calibration 1 - cow) and press the button Enter.The
following message appears on the display:
Calibration 1
Getting ready
6.4.2.2. Entering values for the separate sample parameters
Wait for the sound signal and the following menu on the display:
Cal1 Samp High
Fat=f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit
Operation manual
MRC.VER.01-02.10
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)