1 error messages – KERN & SOHN ABJ 320-4NM User Manual
Page 96
![background image](/manuals/697265/96/background.png)
96
ABS-N_ABJ-NM-BA -e-1312
18.1 Error messages
Error message Explication
Remedy
ERR H
Hardware error
Switch balance off and on
again. If the error message
remains inform manufacturer.
ERR C
High zero point shift during
adjustment
Use ON/OFF to return into
weighing mode. Restart
adjustment process.
Objects present on weighing
plate.
Missing weighing plate
CAL D
Display instable
Check ambient conditions
(draught, vibrations etc.) Use
ON/OFF
to return into weighing
mode. Restart adjustment
process.
ERR N
Error at numerical input (e. g.
wrong password)
Correct input
ERR W
Wrong application
The balance changes to the
previous state. Correct
application.
COM ERR
Wrong remote control order.
The balance changes to the
previous state. Correct remote
control order.
- OL
Missing weighing plate
Install weighing plate correctly
OL
Overload
Reduce load
ABORT
Process cancelled
WAIT
Wait for process
BUSY
When adjustment was started
objects were present on
weighing plate
Remove object and continue
adjustment
Should other error messages occur, switch balance off and then on again. If the error