1 reference weight = 100 – KERN & SOHN ABJ 320-4NM User Manual
Page 61

ABS-N_ABJ-NM-BA-e-1312
61
14.2.1
Reference weight = 100 %
Select percent weighing mode, see chpt. 14.2
Press TARE.
If necessary, press the navigation buttons ( )
repeatedly until „SAMPLE“ is displayed.
Press TARE.
Display, if no reference weight is stored
Display, if a reference weight is stored
Set reference
If necessary, place an empty container on the balance
and tare it by pressing TARE.
Press MENU twice.
Put a reference weight which corresponds to 100 %.
(Minimum weight: Readability d x 100)
Wait for the stability display ( ), then confirm by
pressing TARE.
From here on the weight of the sample is displayed in
percentage in terms of the reference weight, see chpt.
13.2.3.