Confirm with TARE.
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
5. Start tolerance control
If necessary, place an empty container on the balance and tare it by pressing
TARE
.
Place weighed goods and wait until the indicator
,
or
appears. With
the help of the indicator check if the weighed goods are under, inside or over the
default tolerance
Example of entry:
TG.VAL
100.0000 g
LM.VAL
10.0000 g
Weight less than nominal
weight and below the lower
tolerance
flashes slowly
flashes fast
Weight within tolerance [target
value ± tolerance]
(90.0000g – 110.000g)
Weight more than the nominal
weight and above the upper
tolerance
flashes fast
flashes slowly