69

0. Call up menu

In weighing mode press the MENU button twice

1. Select function

Press the navigation buttons ( ) repeatedly until

TOOLS“ is displayed.

Confirm with PRINT

Press the navigation buttons ( ) repeatedly

until „CHECK.W“ is displayed. The current setting is

marked by the stability display ( ).



Further steps depend on the desired setting:

Stability
display (

)

Function

set/update

cancel

yes

ON

Press PRINT and

continue with step 3.

Press TARE and

continue with step 4.

no

off

Press TARE and

continue with step 3.

continue with step 4.

2. Set limit values

When entering the limit values ensure that the values match logically one

with another, i.e. the lower limit value must not be greater than the upper

one.
If not considered, the balance will adapt the limit values automatically.

Press the navigation buttons ( ) repeatedly until

HI.LIM“ is displayed.

Press TARE. The current setting will be displayed.

Use the navigation buttons ( ) to enter the desired

value, see chpt. 3.1.1 „Numeric input“.