3 formula mode – KERN & SOHN ABJ 320-4NM User Manual
Page 64
![background image](/manuals/697265/64/background.png)
64
ABS-N_ABJ-NM-BA -e-1312
14.3 Formula mode
The formula function allows to add on various components of a mixture. For check
purpose, the weight of all components (CMP001, (CMP002 etc.), as well as the total
weight (TOTAL) can be printed out. The balance works with a separated memory for
the weight of the weighing container and of the formula components.
In formula mode the „Auto zero“ function is not active (see chpt. 12.2).
1. Connect printer (see chpt. 15 „Data output“)
2. Call up menu
In weighing mode press the MENU button twice
3. Select formula mode
Press the navigation buttons ( ) repeatedly until
„APL.FUNC“ is displayed.
Confirm with PRINT
Press the navigation buttons ( ) repeatedly
until „FORMULA“ is displayed. The formula symbol
[ ] appears.
Press TARE. The indicator
is displayed, from
here on the balance is in formula mode.
Enable the edition „component no.“(see chpt. 14.3.1)
and „Total“ (see chpt. 14.3.2).
4. Weighing components
If necessary, place an empty container on the balance
and tare it by pressing TARE.
Press PRINT , the symbol
goes out.
When the GLP function (see chpt. 8.3) is activated, the
headline will be printed.
Weigh first component.
Press PRINT.
When standstill control ( ) is completed, the weighing
value of the first component (CMP001) is issued to the
optional printer. The displayed value is added into the
total adding memory. This is followed by the automatic
taring, the symbol [
] is displayed.