beautypg.com

A&D Weighing HC-3Ki User Manual

Page 55

background image

53

14. OP-04 RS-232C & RELAY OUTPUT

RS-232C interface and relay output for comparator results are obtained by installing
OP-04.
‰

The OP-04 unit includes an interface board, a connector plug (DIN type) and two

screws. (M3x6 tapping type).

OP-04 Installation

‰

See “13-1. Installation”.

‰

OP-04 is installed to the same slot with the OP-03.

‰

The RS-232C specifications are same as the OP-03 RS-232C interface and refer to

“13. OP-03 RS-232C SERIAL INTERFACE”.

Comparator Relay Output Circuit

Pin connections















‰

The HC-i is designated as DCE (Data Communication Equipment).

Maximum rating of the Relay Output

The maximum rating of the replay output is as follows.

‰

Maximum voltage:

50V DC

‰

Maximum current:

100mA DC

‰

Maximum ON resistance

30Ω






(View from outside)

1 HI (relay output)
2 Receive data (RS-232C)
3 Transmit data (RS-232C)
4 LO (relay output)
5 Signal ground (RS-232C)
6 OK (relay output)
7 Data set ready (RS-232C)
8 COM (relay common)

Inside of the HC-i

Mating connector: JA+TCP0586
(Included in the OP-04)

This manual is related to the following products: