F-function parameters, 88clear8, Aidal 0 0 00 – A&D Weighing HC-3Ki User Manual
Page 32: Aidal 1 0 00
30
9. F-FUNCTION PARAMETERS
9-1. To Change or View F-Function Settings
Start with the scale switched off.
1. Press and hold the
ZERO
key, then press
the
ON/OFF
key.
The count display will show “f-00” with
“00” blinking. Then release the both keys.
2. Press the
0
Æ
9
keys to display the
number of the F-Function section.
For example: the
5
key to enter into the
F-Function F-05-X Comparator section.
3. Press the
ENTER
key.
The count display will show the
F-Function and its present setting will
blink.
4. You may now either change the setting (Step
5) or move to the next F-Function (Step 6).
5. Use the
0
Æ
9
keys to change the
setting.
The
C
key clears the input setting if you press
the wrong key and want to re-enter.
If you make a mistake and want to escape without saving any changes made after
the last time the
ENTER
key was pressed – press the
ON/OFF
key to exit.
After the
ENTER
key is pressed, the data is entered.
6. Press the
ENTER
key to save any
changes and/or move to next.
7. When finished: press the
ON/OFF
key to
exit. Then, press it to turn the display back ON.
New settings will operate as set.
ON
OFF
ZERO
+
8
f-00
810
8
f-05
810
8
f-05-01
88Clear8
8
aidal
0
0
00
8
f-05-01
88Clear8
8
aidal
1
0
00
ENTER
8
f-05-02
88Clear8
8
aidal
0
0
00
ENTER