beautypg.com

Comparator function, 88clear8, Aidal 0 0 00 – A&D Weighing HC-3Ki User Manual

Page 27: Aidal 1 0 00

background image

25

7. COMPARATOR FUNCTION

‰

The scale contains a comparator function that checks the amount on the weighing

pan against set acceptable count or weight levels. When the comparator function is
activated, “HI”, “OK” or “LO” indicator will be displayed.


‰

Before the comparator will work, Upper

and Lower Limits must be set (see
below). The levels are set by count or
weight. So, if you are using weight for
your comparator levels, calculate the
weight before starting the procedure
below.


‰

If the OP-04 is installed, comparator

relay output is also available.


‰

The comparator responds as follows,

“HI”

Upper Limits < Count / Weight Data

“OK” Lower

Limits

≤ Count / Weight Data ≤ Upper Limits

“LO”

Count / Weight Data < Lower Limits

To Set the Comparator

‰

Start with the scale switched off.


1. Press and hold the

ZERO

key, then press

the

ON/OFF

key – release both.


The count display will show “f-00” with
“00” blinking.


2. Press the

5

key to enter into the

F-Function

F-05-X Comparator

section.


3. Press the

ENTER

key.

The count display will show the
F-Function and its present setting will
blink.


4. Use the

0

Æ

6

keys to display the

number of the desired setting.

For example, let’s select “1” compare all

data.

Upper limit

102 pcs

Lower limit

98 pcs

The beeper is set ON at “OK”.

ON

OFF

ZERO

+

COUNT

SAMPLE

TOO

LIGHT

HI

LO

OK

88888

100

‘

Ñ

beep, beep …ґ

8

f-00

810

5

8

f-05-01

88Clear8

8

aidal

0

0

00

8

f-05-01

88Clear8

8

aidal

1

0

00

ENTER

8

f-05

810

This manual is related to the following products: